7.8.
Installation of the power supply
DANGER
Explosion hazard
Before starting the mounting and installation work, make sure
to have a “hot permit” and monitor at all time the ambient
atmosphere with a certified flammable gas detector.
DANGER
Electric shock.
Electricity may cause death. Only educated staff should be
allowed for mounting and installation job.
100-230 Vac/ 47-60 Hz/ 300W
Fuse 4A slow (by user)
N
L
PE
Pre-fuse
N
L
PE
Cable gland (M16):
For cables with Ø 3,5-10 mm.
Remote power supply circuit
to guarantee power down the
analyser in case of hazard. (See
chapter 7.9)
K2 switch off relay
G5
G5
2
~ ~