MRU GmbH
SWG 100 BIO Ex User Manual
Page
42
of
150
7.6.
Correct tubing of the SWG100BIO-Ex
The analyzer uses two different fitting diameters. The position and the diameters of the
different connectors will be shown in the sketch below:
DANGER
Explosion hazard
Before starting the mounting and installation work, make sure to
have a “hot permit” and monitor at all time the ambient
atmosphere with a certified flammable gas detector.
CAUTION
Do not cover the breather drain!
NOTE
All connectors of the analyzer use female threads. The analyzer may
be connected by tubing or piping.
Distance, which should be fulfilled
Top-side
On the top of the analyzer are the sample gas-inlets. Sample gas inlet 1 is the default gas
inlet port. The analyzer may be equipped with up to 3 additional sample gas inlet ports
from different sampling sites. The following sketch shows the position of the different
sample gas inlets.
Min. 40 cm
Min. 30 cm
Min. 50 cm
Do not cover the
breather drain.