
MRU GmbH
SWG 100 BIO Ex User Manual
Page
41
of
150
For outdoor installation
Ensure that the analyzer is mounted on a rain and sun protected place (weather shade).
For indoor installation:
Ensure that the analyzer is installed on a dry and clean place. Be sure that the room is
permanently vented with fresh air (forced ventilation).
Connect the VENT gas-outlet of the device to ambient by using adequate tube with min
Ø8mm ID.
The device must be mounted in a room, which is ventilated enough. The vent must tube
above into an area, where is no ignition source.
In a radius of min. 30 cm the vented gas can have the potential to be explosive.
DANGER
Explosion hazard
Be sure, that around the vent is no ignition source.
Every sample point must be equipped with a safety valve, which allows the operator to
close the sample point from the analyser quickly.
DANGER
Safety valve
The operator must install a safety valve on every single sample
point. The valve must be resistance against the sample gas.
Max. x6 gas inlets
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...