Zeroing (Cycle phase details):
In the cycle phase detail “
Z
EROING
” the zeroing time
can be configured.
ZEROING
Measuring site valves
Valves closed
Zeroing valve
Valve open
Duration
2min to 1 h
Recommendation
5min., in general not to be changed by user as
depending only on analyzer internal setup
Measurement SPX (Cycle phase details):
In the cycle phase details of “Sample
point X” the measurement time and the suction delay can be configured. Each
sample point can be configured individually. In the cycle phase details the
following times can be set:
MEASUREMENT SPx
Measuring site valves
Valve of selected site is open, others closed
Zeroing valve
Valve closed
Duration
Phase duration: 2 min. to 24 h
Suction/response time: 30 sec. to 1h
Pure measurement: calculated
H2S-low: Activated/protect (Optional)
Stand-by (cycle phase details):
In the cycle phase details “Stand-by” the sleep
mode time can be configured. In the cycle phase details the following times can be
set:
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...