6.
Sample Gas measurement
6.1.
Cycle configuration
Path and default setting
E
XTRA
/
M
EASUREMENT CYCLE CONFIG
.
When the menu
“M
EASUREMENT CYCLE CONFIG
.”
is selected the user definable setting for
the measurement cycle will appear (see screenshot below).
General information
The menu point
“
CYCLE
CONFIGURATION”
allows the user to configure an individual
measurement cycle. Every installed sample point can be configured. For the
configuration the user has the following phases, which can be selected:
•
Zeroing.
•
Purging.
•
Stand-by.
•
Measurement SPx (SPx stands for Sample point 1, 2…).
The configuration is performed with the three function keys
F1
,
F2
and
F3
.
•
F1
:
Delete a phase.
•
F2
:
Make an Auto-config.
•
F3
:
Insert a new phase.
•
OK
:
View/change phase details
•
Left/right
:
Change the phase type.
Auto configuration
With
F2
the “Auto-config.” can be selected. The user can select one of two default
cycle configurations.
•
One zeroing / cycle.
•
One zeroing / sample point.
The first program is for applications where the different measurement points have
almost the same gas concentrations. The zeroing is not necessary at every change
of the measurement SPX.
The second program is for applications where the different measurement points
have different gas concentrations. A zeroing is recommended after every
measurement point change.
The screenshots below show the “One zeroing / cycle” and “One zeroing / sample
point” in comparison.
Screenshot shows default setting,
when the
„M
EASUREMENT CYCLE CONFIG
.”
will be started the first time.
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...