MRU GmbH
SWG 100 BIO Ex User Manual
Page
143
of
150
18.1.
Analyser Status (address 0 and some mirror addresses)
The Analyser Status is a 32bit-word and must be interpreted bitwise.
Bit
Description
0
Power-On
(until the first zeroing has been done)
1
System-Alarm, see table below
2
Air Purging (zeroing)
3
Gas Sampling (preparing measurement, not measurement!)
4-7
Currently sampled sample point number
(1...4, reads 0 while air purging)
8
One sensor (e.g. CO or H2S) is currently purged
9
One sensor (e.g. CO or H2S) is currently cut-off the sample gas
10-27
reserved for later applications
(reads zero)
28-31
External Control's state, range 0…15
Some status examples:
Decimal
Hexadecimal
Binary
state description
1
0001h
0000 0001
Power-On (self-test)
5
0005h
0000 0101
First Air Purging (Power-On + Air Purging)
24
0018h
0001 1000
Preparing meas. smp.pt.1 (Gas Sa
smp.pt.1)
16
0010h
0001 0000
Measuring sample point 1
24
0018h
0001 1000
Preparing meas. smp.pt.2 (Gas Sa
smp.pt.2)
32
0020h
0010 0000
Measuring sample point 2
48
0030h
0011 0000
Measuring sample point 3
18
0012h
0001 0010
Measuring sample point 1 + System-
Alarm
0
0000h
0000 0000
Adjustment (user interaction at the
analyser)
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...