Position 4: PTFE Filter (#51513)
Required materials: PTFE filter (#51513) contained in the service set.
Required tools: needle-nose pliers.
Exchange interval: Exchange necessary, if gas flow is too low.
Steps:
Pull the exhausted filter-unit from the viton
tube and exchange it with a new one. If it
necessary, the tube can be solved from the
tube with a needle-nose pliers.
Position 5: Sinter filters (#65988)
Required materials: Sinter filters (65988), contained in the service-set.
Required tools: Spanner (range:17)
Exchange interval: Every 2 month, dependent from the dust- and pollution load.
Steps:
1.
Remove the gas tube. Be sure, that no
process gas emits in the environment.
The upper part from the nozzle can be
removed with a spanner (range 17).
2.
Inside of the nozzle is a sinter filter
tablet with an o-ring. Remove the
exhausted sinter filter tablet and the
o-ring and exchange both parts to a
new one. The compensation parts can
be found in the service-set.
3.
The gas tube can now connect with
the nozzle again.
O-ring (#64798)
Sinterfilter
(#65988)
Copper seal
(#61947)
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...