MRU GmbH
SWG 100 BIO Ex User Manual
Page
115
of
150
Review
Recommended
intervals
Actions
Moisture in the analyzer.
Weekly
Remove the moisture. Call
vendor specialist.
Dirt and depositions in analyzer.
Weekly
Remove the dirt, prevent further
penetration of dirt.
Inform vendor.
Dirt and moisture in the filter- unit
Weekly
Exchange the filter-unit.
Testing the gas pipes of leakage
with “sniffer”.
Weekly
Tighten or exchange the gas
pipe if it is necessary
Inspecting the conditions of the gas
filters
Every month
Exchange if is necessary
Cleaning the surface and window of
the analyzer, to prevent a
electrostatic loading of the cabinet.
Weekly.
Clean the surface and the
window with an antistatic cloth.
Check the switch-off relay and the
cabinet ambient solenoid valve.
Every month
See chapter 9.2.
Look if all flexible tubes are tight.
Exchange flexible tubes, which are
exhausted or not tight enough
Weekly
Exchange the tubes to new one.
Exchange all peristaltic hoses,
unrequired if the hoses are defect or
not.
6 month
Exchange all peristaltic hoses,
unrequired if the hoses are
defect or not.
In the follow chapters, there will be introduced some service parts, which are important
for the reliable operation. These parts are independent from the regular checks and
must be replaced in a regular interval of minimum 6 month.
Summary of Contents for SWG100 BIOEX
Page 1: ...9512EN BIOEX SWG100 BIO EX USER MANUAL...
Page 2: ...MRU GmbH SWG 100 BIO Ex User Manual Page 2 of 150...
Page 30: ......
Page 47: ...MRU GmbH SWG 100 BIO Ex User Manual Page 47 of 150...
Page 48: ......
Page 98: ......
Page 106: ......
Page 122: ......
Page 147: ...MRU GmbH SWG 100 BIO Ex User Manual Page 147 of 150 20 Declaration of conformity...
Page 148: ...MRU GmbH SWG 100 BIO Ex User Manual Page 148 of 150...
Page 149: ...MRU GmbH SWG 100 BIO Ex User Manual Page 149 of 150...
Page 150: ...MRU GmbH SWG 100 BIO Ex User Manual Page 150 of 150 EN_V3 0_BIO EX...