3.2.
Select measurement point
The measuring point should be in an area where the flow no longer has any
particular turbulence. That is:
•
Not directly before or after a pipe elbow. The measuring point
must be at least 3 times the chimney diameter from a bend.
•
The probe may be mounted both horizontally and vertically.
•
The opening should have access to the core flow. This is usually
the centre of the pipe.
•
For installations on a metal pipe, make sure that the wall
thickness can support the weight of the probe. For this purpose
the mounting flange is welded to the pipe.
•
For installation on a brick fireplace, it is advantageous to screw
a steel plate with the correct hole diameter to the brick
fireplace.
3.3.
Installation of the analyzer
The analyzer is equipped with adapters for the wall mounting. The dimensions
for the wall mounting are shown in picture below.
Above 3 times of diameter
Over 3 times of diameter
Core flow
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...