USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
9 /
117
2.1.
Intended use
The analyser MGAprime Q is proofed according to DIN EN 15267-4.
The following table lists proofed and non-proofed measured values.
Measured variables
and conversions.
Proofed according to
15267-4
Notes
O
2
CO
NO
NO
2
CO
2
SO
2
N
2
O
Flue gas temperature -
Air temperature
-
Measurement of the
ambient temperature
Draft / Differential
Pressure
-
Velocity (v)
-
Measurement with
Pitot tube
ppm -> mg/Nm³
-
NOx: mg/Nm3 NO
2
-
O
2
Ref
-
mg/kWh
-
The proof according to DIN EN 15267-4 was performed under the
following conditions:
Grip heating temperature
160 °C
Heating hose temperature
160 °C
3m Heating hose
Zero-point interval
deactivated
The following functional modules are not part of the proof according to
DIN EN 15267-4:
•
Data Storage
•
Internal logger