USER MANAUL
MGA
prime Q
MRU GmbH, D-74172 Neckarsulm
105 /
117
►
Press
A window appears.
Start drying manually
If the "condensate self-help" was not successful and there is still water in
the gas duct, you must start a drying of the gas duct manually.
A corresponding message is displayed and an acoustic signal sounds.
CAUTION
Acid from the condensate
Acid burns may result from weakly acidic liquids from the
condensate.
►
If you come into contact with acid, wash the area immedi-
ately using a lot of water.
►
Note the safety data sheet for phosphoric acid (10%).
NOTE
►
Dispose of acidic condensate and phosphoric acid in
accordance with national and, if necessary, local dis-
posal guidelines.
►
Remove the round filter.
►
Dry the round filter, if necessary.
►
Press OK.