18.2.
Condensate alarm
A condensate alarm occurs as soon as water is detected in the gas duct.
In this case, the measurement can be implausible.
A condensate alarm that has occurred can be eliminated in a two-step
process.
After a condensate alarm occurs, the analyser attempts to remedy the
problem independently by means of a "condensate self-help" and to
switch back to measuring mode
If the "condensate self-help" was not successful and there is still water in
the gas duct, you must start a drying of the gas duct manually.
Condensate self help
With "condensate self-help" you do not have to actively initiate any
steps. The measuring analyser carries out the "condensate self-help" in-
dependently.
As soon as water is detected in the gas duct, the analyser displays a
message.
The
“Condensate
self-
help”
is performed.
As soon as the "Condensate self-help" has been successfully com-
pleted, a message appears.
A zero point measurement is started automatically.
You can display the start and end of the "Condensate self-test