Possible reasons the test has failed.
Condensate Separator:
Defect or missing O-
Rings / cracked
condensate show glass
Probe handle and probe tube:
Defect or missing O-Rings.
Or the connection is not tight enough.
Hoses:
Defect hose - very common at
the condensate separator port.
Hoses:
Defect hose - very common at the probe handle.
If you have this PRE-Filter (# 56356) installed, check the show glass
for cracks and check the O-Rings
Also possible:
Internal pump is not working or doesn’t have enough suction.
Spare parts.
POS.
Parts for Condensate Separator
Part #
POS.
Parts for Probe
Part #
12
Filter housing with O-Ring
63919
9
O-Ring 8 x 2.4 (probe)
61065
4
O-Ring 22 x 1.5
64120
10
O-Ring 3.7 x 1.9 (probe)
59582
6
Show glass
62627
NA
O-Ring 24 x 1.5 (PRE-Filter #56356)
51516
9
O-Ring 5 x 1.5
61464
NA
Show glass (PRE-Filter #56356)
52648
NA
Viton sampling line
55641
POS.
Parts for Analyzer
Part #
NA
O-Ring 4 x 1.5 (draft port)
61463