26
Adjust Ratio
Due to long time usage of device, the sensitivity of IR detector
may decrease. In such case, it is necessary to adjust ratio value
to get more accurate temperature reading. During calibration, if
temperature reading is higher than actual target temperature,
change ratio value lower, and vise versa. Press
▲
or
▼
once to
change ratio value 0.01. Long press
▲
or
▼
will change ratio
value in 0.1 intervals.
Adjust Temp
Under certain condition, it is necessary to adjust temperature
reading by a pre-define value. Normally this option should be left
unchanged. Press
▲
or
▼
once to change 0.1
O
C. Long press
▲
or
▼
will change this value in 1
O
C interval.
Save Setting
All parameter changes must be saved before power off device.
Otherwise, changes will not be saved for next power on.
5.2.3 Image
There are 5 items: Image Setup, Analysis Setup, Image Reverse,
Clear Screen, and Auto/Manual (L&S) in Image menu.