CURRENT,VOLTAGE,Type K 
CALIBRATOR

Model : CC-423

Your purchase of this 3 in 1
CALIBRATOR  METER
with Source & Measurement
marks a step forward for you
into the field of
precision measurement.
Although this meter a
complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.

 OPERATION MANUAL

PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 

3, Hagavish st. Israel 58817  Tel: 972 3 5595252, Fax: 972 3 5594529   

[email protected]

 

MRC.

1

.1

8

 

Summary of Contents for CC-423

Page 1: ...gh this meter a complex and delicate instrument its durable structure will allow many years of use if proper operating techniques are developed Please read the following instructions carefully and always keep this manual within easy reach OPERATION MANUAL PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION 3 Hagavish st Israel 58817 Tel 972 3 5595252 Fax 972 3 5594529 mrc mrclab com MRC 1 18 ...

Page 2: ...VIDE PROCEDURE 6 4 3 Type K SOURCE PROVIDE PROCEDURE 7 5 MEASUREMENT PROCEDURE 7 5 1 CURRENT MEASURE PROCEDURE 7 5 2 Power and current measurement of two wire loop Procedure 7 5 2 Type k MEASURE PROCEDURE 8 6 ADVANCED SETTING 8 6 1 Auto power OFF management 9 6 2 Select the Temp unit to or 9 7 POWER SUPPLY from DC ADAPTER 10 8 BATTERY REPLACEMENT 10 9 SYSTEM RESET 10 ...

Page 3: ... up to 500 ohms Four function provide the quality and accuracy of handheld calibrator of precision current source 2 SPECIFICATIONS 2 1 General Specifications Circuit Custom one chip of microprocessor LSI circuit Display LCD size 52 mm x 30 mm LCD with white backlight ON OFF Measurement Current Max 24 mA Type k Temp 199 9 to 1370 0 Source Current 0 to 24 00mA Voltage DC mV 200 0 to 200 0 mV Type k ...

Page 4: ...urrent source Approx DC 18 mA LCD Backlight is ON Approx DC 5 mA Current source under 10 mA signal out put Approx DC 33 mA Backlight is ON Approx DC 5 mA Weight Meter 306 g 0 67 LB with cable Dimension Meter 178 x 68 x 45 mm Accessories Instruction manual 1 PC Included Hard carrying case CA 06 1 PC V A cable LN TL421 1 PC Type K source cable AC DUK02 1 PC Optional Accessories AC to DC 9V adapter T...

Page 5: ... source Range Display Accuracy Resolution 199 9 mV to 199 9 mV 0 1 mV 0 25 FS 0 2mV Output measured load inpedence should 1 K ohms Type k source Type k source Range Display Accuracy Resolution 199 9 to 1370 0 0 1 degree 100 0 1 FS 1 100 0 1 FS 2 327 0 to 2498 0 0 1 degree 148 0 1 FS 2 148 0 1 FS 4 intput inpedence 10 12 ohms Remark The above specification are tested under the environment RF Field ...

Page 6: ...e Type k measure Type k measure Range Display Accuracy Resolution 199 9 to 1370 0 0 1 degree 0 1 FS 1 0 327 0 to 2498 0 0 1 degree 0 1 FS 2 0 intput inpedence 10 12 ohms Remark The above specification are tested under the environment RF Field Strength less than 3 V M frequency less than the 30 MHz only CAUTION Do not apply a voltage exceeding the maximum input voltage otherwise the input part may ...

Page 7: ...ery Cover Screws 3 3 Power Button 3 13 Tripod Fix Nut 3 4 button 3 14 Stand 3 5 Button 3 15 Battery Compartment Cover 3 6 Button 3 16 Type K Temp Input Socket 3 7 FUNCTION Button 3 17 Cable 4 DIN Socket 3 8 Source Measure Button 3 18 Cable 4 DIN Plug 3 10 Reset Button 3 19 Red Alligator clip 3 20 Black Alligator clip 5 ...

Page 8: ...urrent output same as the display value Remark The Red Alligator clip 3 19 Fig 1 is for the positive current output The Black Alligator clip 3 20 Fig 1 is the ground for current output 4 2 VOLTAGE SOURCE PROVIDE PROCEDURE 1 Power on the meter by press the Power button 3 3 Fig 1 2 sec 2 Press the Source Measure Button 3 8 Fig 1 select to SOURCE Function 3 Press the FUNCTION button 3 7 Fig 1 select ...

Page 9: ...to rapidly adjust the value And will generate the Type k Temp output same as the display value 5 MEASUREMENT PROCEDURE 5 1 CURRENT MEASURE PROCEDURE 1 Power on the meter by press the Power button 3 3 Fig 1 2 sec 2 Press the Source Measure Button 3 8 Fig 1 select to MEASURE Model 3 Press the FUNCTION button 3 7 Fig 1 select to CURRENT Measure Function 4 The Red Alligator clip 3 19 Fig 1 connect to ...

Page 10: ...urement input The Black Alligator clip 3 20 Fig 1 is for the negative current measurement input b The measuring procedures principal are same as the 5 1 Current measurement except the meter build in the DC 12 V power source in serial with the current meter please refer Fig 2 5 3 Type k MEASURE PROCEDURE 1 Power on the meter by press the Power button 3 3 Fig 1 2 sec 2 Press the Source Measure Butto...

Page 11: ...turn to normal screen 6 1 Auto power OFF management When the lower display show PoFF 1 Use the Button 3 5 Fig 1 or Button 3 6 Fig 1 to select the upper value to yES or no yES Auto Power Off management will enable no Auto Power Off management will disable 2 After select the upper text to yES or no press the Source Measure Button 3 8 Fig 1 will save the setting function with default 6 2 Select the T...

Page 12: ...ent become inaccurate 2 Loose the screws of the Battery Cover Screws 3 12 Fig 1 and take away the Battery Cover 3 15 Fig 1 from the instrument and remove the battery 3 Replace with DC 1 5 V battery UM3 AA Alkaline heavy duty x 6 PCs and reinstate the cover 4 Make sure the battery cover is secured after changing the battery 9 SYSTEM RESET If the meter happen the troubles such as CPU system is hold ...