26
8.
Calibration
Calibration
should
take
place
at
20-25℃,
kept
constant
within
±0.5℃.
A
dedicated
calibration
software
will
write
calibration
values
in
your
BDI-DIG
,
after
the
distilled
water
has
been
repeatedly
weighed
up
at
least
five
times.
Hardware
needed::
Electronic balance with readability of 0.01 mg
Distilled water
X86-or X64-architectured PC with pre-loaded
Windows(XP/Vista / 7/8/10 )operating system
Software needed:
Dedicated
calibration
software
of
BDI-DIG
(For more information, please contact with your nearest
distributor.)
CAUTION:
If
your
BDI-DIG
can
not
work
properly
after
calibration,
please
contact
your
nearest
distributor