Connecting the PML²
9
Important! Do not connect Waste Fluid Hose to a pressurised system, as this
will cause the Analyser to malfunction and could cause internal leakage. The
Waste Fluid Hose must be discharged into a waste disposal bottle, or into a
tank/vessel vented to atmosphere if running a Continuous test due to the large
quantity of oil discharged during a test.
4. Connect
WASTE FLUID HOSE
to Analyser (waste connector). Push back quick
coupling outer ring before connecting / disconnecting hose end.
5. Connect
FLUID SAMPLING HOSE
to Analyser (HP connector)
6. Connect
FLUID SAMPLING HOSE
to the system by means of the minimess con-
nector.
The system to be monitored must not exceed 400 bar or be less than 2 bar.
7. Installing
LPA-View Software
– refer to the LPA-View User Manual and follow the
CD Installation instructions.
8. Connect
Multi-core cable
into the PC and the other end into the small connector
on the PML². If the PC has only USB ports, use the USB to serial converter and
follow the instructions on page 12
9. 500 µm Coarse Screen Filter (Optional Equipment) Screw this filter directly onto
the PML² minimess connector, and then connect the minimess hose to the other
end. This filter is recommended for heavily contaminated systems.
3.1 Methods of Operation
Users may operate the PML² in two ways:-
1. PC Operation
2. Remote Display (optional equipment) see page 30