MOXTEK X-ray Sources MOX-HPC150 Max Operation Manual Download Page 1

MOXTEK

®

TUB-MAN-1020 Rev C

Subject to technical change without notice

MOX-HPC150 Max

Operation Manual

Table of Contents

•  Introduction
•  X-ray Source Characteristics
•  Tube Current and Power Operational Range
•  Initial Inspection
•  Handling
•  Mechanical
•  Tube Setup and Operating Instructions
•  Operating Precautions - Warnings

1
2
2
2
2
3
4
7

1

X-ray

 

Sources

Introduction

Moxtek HPC 150W 75kV X-ray tubes are designed to be used in XRF 

and imaging applications.  The tube is designed to be air-cooled and 

the cooling can be adapted to meet specific needs.  The push/pull fan 

configuration makes it easy to direct the warm air outward.  For XRF 

applications the cone is designed for a vacuum seal at its base.  The HPC 

tube incorporates radiation shielding.

Figure 1 - The Moxtek HPC150

Summary of Contents for X-ray Sources MOX-HPC150 Max

Page 1: ...Operating Precautions Warnings 1 2 2 2 2 3 4 7 1 X ray Sources Introduction Moxtek HPC 150W 75kV X ray tubes are designed to be used in XRF and imaging applications The tube is designed to be air cool...

Page 2: ...um Power Long Life Power Maximum mA Long Life mA Tube Current and Power Operational Range When a Mox HPC tube is received it should be unpacked and inspected as soon as possible Caution Do not lift or...

Page 3: ...ken when handling the tube Do not drop or strike the tube on any surface Care must be taken to not damage the beryllium window or high voltage cable Support the weight of the tube when lifting or carr...

Page 4: ...ply by connecting the red lead to the positve and black lead to the negative terminals on the power supply The power supply should provide the correct power watts for two fans Caution Reversing these...

Page 5: ...ions cont 4 Install the cable collar onto the tube cable if it is not installed already Figure 9 Cable collar 5 Clean the cable connector with a lint free wipe 6 Apply a small amount of silicone oil a...

Page 6: ...0 2 mA steps the first time the tube is operated The maximum power allowed is 150 watts and voltages up to 60kV within the provided power and volatge combinations Caution Do not exceed 150 watts of tu...

Page 7: ...If the Beryllium target is broken a very small amount of dust might be generated If this occurs follow the instructions on the MSDS for cleanup MoxHPC X ray tubes may become very hot during operation...