22
7 Usage
Instructions for Use 2019-01-16
7.3 Basic Operation
7.3.1
Turning Power On/Off
On the right side of the unit, turn the Power
Switch on ( ).
The Power Lamp lights up in green.
After use, turn the Power Switch off (
)
The Power Lamp will go out.
Be sure to check the following points before using the Lubrina 2.
• Is the air pressure setting between 0.3 MPa and 0.5 MPa? (Recommendation: 0.4 MPa)
• Does each coupling have proper O-rings?
• Are all connection nuts for couplings properly secured?
1. Open the front door.
• Never move your face close to the unit during the chuck lubrication. Otherwise, oil might
spray out into your eyes.
7.3.2 Chuck Lubrication
On
Off
Power Lamp
Power Switch
Hold up
approx. 4 sec.
Nozzle
Nozzle Skirt
Pay attention when lifting the handpiece up. Do not tilt the handpiece against the chuck
nozzle. Doing so could damage the nozzle.
If the handpiece is removed from the chuck nozzle before the single sustained beep, it
may result in scattering a huge amount of oil.
If the nozzle skirt comes off, be sure to reinstall it again. Otherwise, oil might spray out
into your eyes.
If the device has not been used for some time, be sure to
check that it works properly before using it again.
2. Line up the nozzle tip of the chuck lubrication unit with the hand-
piece bur insertion hole, and then insert the nozzle into the hand-
piece chuck.
Lift the handpiece straight up to start chuck lubrication.
Hold it up for approximately four seconds until you hear a single
sustained beep.
If you hear a series of beeps instead of a single sustained beep, try
again.
Chuck Lubrication Unit
•
Be sure to turn the Power Switch off after use.
• If the device will not be used for a while, unplug it and close
the main air valve.
* Chuck lubrication is dedicated to air turbine handpieces and contra
angle handpieces that use FG burs.