Instruction Manual PM-5947 Copyright © 2022 

 

 

C705P Digital Indicator Meter  

Instruction Manual 

 
 

 

 
 
 
 
 

Summary of Contents for C705P

Page 1: ...10 20 2022 Rev 1 Instruction Manual PM 5947 Copyright 2022 C705P Digital Indicator Meter Instruction Manual ...

Page 2: ...1 10 General G Menu Descriptions 11 1 11 Entering the User COM1 A1 Menu 12 1 12 Leaving the User COM1 A1 Configuration Menu 13 1 13 User COM1 A1 Menu Descriptions 14 1 14 Entering the COM2 A2 Menu 14 1 15 Leaving the COM2 A2 Configuration Menu 15 1 16 COM2 A2 Menu Descriptions 15 INSTRUMENT CALIBRATION 16 2 1 Calibration Overview 16 2 2 Live Calibration Overview 16 2 3 Live Span Calibration Instru...

Page 3: ...hows T The Compression mode shows mV V Indicates that the unit of the displayed value is in millivolts per volt Tf Indicates that the unit of the displayed force is in tonne force klbf Indicates that the unit of the displayed force is in kilopounds kips lbf Indicates that the unit of the displayed force is in pounds kN Indicates that the unit of the displayed force is in kilonewtons N Indicates th...

Page 4: ...lbf kgf and N If Primary Unit is klbf Tf or kN then avail units are mV V klbf Tf and kN Zero The Zero key functions as a Tare 0 on display Peak Freeze Peak Selects the peak hold display mode of the instrument e g live data positive peak and negative peak Freeze Pressing this key causes the display to stay constant stop updating until this button is pressed again Profile Selects the one of the five...

Page 5: ... bracket to a wall or table CONNECTIONS The rear cover must first be removed to make the appropriate connections to the weigh platform etc To remove the rear cover simply remove the screws that secure it to the enclosure and set aside Caution Disconnect power source from instrument prior to removing rear cover Caution Disconnect leads from rechargeable battery if installed to avoid shorts Force se...

Page 6: ... of the force sensor shield wire connection GND Power Connections AC version The C705P instrument ships with a pre installed AC line cord It has been pre wired to Terminal Block J1 at the factory Simply plug the unit into a standard wall outlet USB Port The external USB port is used for communication and updating of the firmware Consult factory for use Getting Started 1 Press and hold the ON key o...

Page 7: ...5 User2 A2 Menu Configures COM2 communication parameters The configuration menus are laid out in the following vertical arrangement Top Menu selection level Note some menus may have a secondary top level e g F and F PF 1 Parameter level Selection level or function level e g span calibration Please review the following chart to get a feel for how to navigate among the various menus and parameters N...

Page 8: ...ample to go from F PF 1 to F PF 2 press the DATA key To go from F PF 2 back to F PF 1 press the PROFILE key 7 Scroll down using the ZERO down key to reach the parameter level The instrument shows PF1 F 1 8 Move from one PF1 F menu parameter to the next by using the PROFILE left or DATA right keys For example to go from PF1 F 1 to PF1 F 10 press the DATA key To go from F PF1 F 10 back to PF1 F 1 pr...

Page 9: ...ress ZERO key to begin sequence PF1 F17 Span Calibration Positive Places instrument into live tension positive span calibration mode Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence 1 5 Entering the Polynomial P Configuration Menu Note This configuration menu has five 5 secondary top levels PF 1 to PF 5 corresponding to the give 5 profiles PF 1 to PF...

Page 10: ...on Menu 1 While on any level except the selection level press the MENU OFF key 1 7 Polynomial Setup P Menu Descriptions This section provides more detailed descriptions of the selections found in the Polynomial Menu Chart The menus are identical for each profile PF 1 to PF 5 Detailed instructions can be found in the next section Instrument Calibration Factory set defaults are shown in bold with a ...

Page 11: ... are identical for each profile 1 Switch off the instrument by pressing the MENU OFF key oFF followed by the PEAK FREEZE key 2 Press and hold the ON key until the instrument beeps and starts to boot up 3 During the countdown phase 555555 444444 etc press and hold the MENU OFF key until the Set message is displayed 4 Press the PEAK FREEZE button while Set is being displayed The instrument displays ...

Page 12: ...ration Factory set defaults are shown in bold with a checkmark CODE NAME DESCRIPTION SELECTION LIST PF1 G3 Calibration Equation Switchover Threshold Selects the range within which the instrument will automatically switch between tension and compression modes 0 0 00000 mV V 1 0 00001 mV V 2 0 00002 mV V 5 0 00005 mV V 10 0 00010 mV V 20 0 00020 mV V 50 0 00050 mV V 100 0 00100 mV V 200 0 00200 mV V...

Page 13: ...ZERO key to begin sequence G21 Factory Reset This sub menu will reset all setup parameters to the default settings It will not overwrite any previously saved calibration data Password protected FACTORY USE ONLY Press the ZERO key twice to execute G30 Special Application Used to select a special application feature 0 None Normal operation 5 Peak Operation 0 5 1 11 Entering the User COM1 A1 Menu 1 S...

Page 14: ...User COM1 A1 Menu Descriptions This section provides more detailed descriptions of the selections found in the User COM1 Menu Chart Factory set defaults are shown in bold with a checkmark CODE NAME DESCRIPTION SELECTION LIST A1 1 Baud Rate Selects the baud rate for data transmission through the serial port 1200 2400 4800 9600 19200 38400 57600 115200 A1 2 Data Parity Bits Selects the number of dat...

Page 15: ...below End test manually by pressing the PEAK FREEZE Set key Key Keycode Menu Off 12 Units 1 Zero 2 Peak EXIT Profile 4 Data 5 1 14 Entering the COM2 A2 Menu 1 Switch off the instrument by pressing the MENU OFF key oFF followed by the PEAK FREEZE key 2 Press and hold the ON key until the instrument beeps and starts to boot up 3 During the countdown phase 555555 444444 etc press and hold the MENU OF...

Page 16: ...on level press the MENU OFF key 1 16 COM2 A2 Menu Descriptions This section provides more detailed descriptions of the selections found in the COM2 Menu Chart Factory set defaults are shown in bold with a checkmark CODE NAME DESCRIPTION SELECTION LIST A2 1 Baud Rate Selects the baud rate for data transmission through the serial port 1200 2400 4800 9600 19200 38400 57600 115200 A2 2 Data Parity Bit...

Page 17: ...1 2 2 Live Calibration Overview Live calibration comprises two main steps zero calibration PFx F16 and span calibration PFx F17 However a third calibration for negative span PFx F15 is also available for systems operating in both tension and compression Here is the recommended sequence for multiple fixtures and multiple calibration points 1 Put force sensor onto test fixture 2 Go to PFx F17 and pr...

Page 18: ...r Err0 The calibration test load or the keyed in load is larger than the full capacity of the instrument Change the calibration test load or check the input data Err1 The calibration test load or the keyed in load is smaller than 1 of the full capacity of the instrument Change the calibration test load or check the input data Err2 There is not enough signal from the force sensor to complete the ca...

Page 19: ...e zero point value The display will show SET and EndC0 momentarily and then revert up to PFx F16 2 6 Entering B coefficient values for Compression Negative Span PFx CBx The B coefficient values for Compression Negative Span are entered one at a time in the P menu using two separate screens pages This example sets CB0 B0 Compression Negative Span to 1 210034E 03 using the PF1 CB0 menu All other men...

Page 20: ...etting of your choice using your calibrator C1 1 While in the Setup mode scroll to G17 and then scroll down once using the ZERO down key The instrument will briefly display Code and then prompt you to enter the password Six zeros are now displayed with one blinking 2 Use the four directional keys to adjust the displayed value to the password and press the PEAK set key The display will momentarily ...

Page 21: ...gnized command from the host device Date is also transmitted when the DATA key is pressed OUTPUT STRINGS TRANSCELL DATA FORMAT String Format 1 is designed for two way communication RECOGNIZED HOST COMMANDS ASCII code Hex Symbol Action by the instrument 50 P Transmit the displayed force data through the serial port 70 p Transmit the displayed force data through the serial port 5A Z Zero the instrum...