PCA Mode

Programming the Pump

  •  Turn the pump on and wait for the self-test to complete.

  •  Highlight 

PROGRAM

 then press 

YES

.

  •  Arrow down to 

NEW PROGRAM

 then press 

YES

.

    (Enter code if asked)

  •  Answer any on screen questions. 

  •  With Unidentified highlighted, press 

YES

.

  •  Select units: Press 

YES

 to 

mg

, unless 

mcg

 or 

mL

 is  

    desired. Then press 

NO

 until desired unit is reached,  

    then press 

YES

.

  •  Enter 

Concentration of drug

, when mg or mcg units  

    are selected, then press 

YES

.

  •  Enter Admin Route: Press 

YES

 for 

IV

, or to change to  

   

Epi

 or 

SQ

, press 

NO

 until desired route is reached,  

    then press 

YES

 to accept.   

  •  Enter 

Load dose

 if ordered, and then press 

YES

.

    (If none then press 

YES

 to 

0.0mL

.)

  •  Press 

YES

 to 

MedLMTS

.

  •  At 

NEXT?

 Press 

YES

.

  •  Enter 

Bag volume

, and then press 

YES

.

  •  Enter 

Basal rate

 (continuous) then press 

YES

.

  •  Enter 

Pt. Bolus

 then press 

YES

.

  •  Enter 

Bolus interval

 (lockout) then press 

YES

.

  •  Enter 

# of Boluses/hr

 then press 

YES

.

  •  At

 Done?

 press 

YES

.

Press 

NO

 to Review the Program.

Press 

YES

 to confirm each setting.

(Review the program prior to Priming.)

Priming the line

  •  At the RUN/OPTIONS screen, press the 

Prime key

  

    and release.

  •  Press 

Prime key

 and Hold.

  •  Release the 

Prime key

 when priming is completed.

  •  Press 

YES

 to exit priming.

Changing an existing program.

  •  Press the 

Pause

 key.

  •  Press 

YES

 to Resume.

  •  Press 

NO

 to Review.

  •  Press 

YES

 to confirm all settings until field to be  

    changed is reached.

  •  At the field to change, enter the new value, then press  

   

YES

 to accept the change.

  •  Press 

YES

 to all other highlighted areas if no other  

    changes are to be made.

  •  At 

Done?

 press 

YES

.

  •  Press 

RUN

 to restart.

Changing the Solution Container

  •  Press 

PAUSE

 key.

  •  Arrow down to 

REPEAT Rx

, press 

YES

.

  •  Answer any on screen questions.

  •  The Pump will auto scroll through the previous program.

  •  Press 

NO

 to Review the program as needed.

  •  Change the solution container.

  •  Press 

RUN

 to start.

Options Menu

  •  LOCK

  •  Clin Dose

  •  SHFT TOT

  •  HRLY TOT

  •  CLR HRLY

  •  DN Occlu

  •  AIR SENS

  •  AUDIO

  •  POWER CK?