Remove thumbscrews and then remove side covers.
2
Mount Cryostation to the optical table using bolt pattern depicted.
3
Fasten screws to a recommended torque value of 100 in-lbs. (11.29 N-m).
1
Locate the red C-spacers on either side of the connection point between the sample
chamber and cryocooler (two total). Unscrew the M3 screws.
2
Remove the red caps then pull out the red tab and remove C-ring.
3
Locate the red shipping support brackets . Using a 5mm Allen key, remove the M6 x
14mm screws on the top of each bracket. Store these screws and washers in the
accessory kit.
4
On the side of the shipping support, locate the two M5 screws holding the bracket in
place. Using a 4mm Allen key, loosen these screws but do not remove them.
5
Slide the shipping bracket (red) up so it is no longer touching the silver tri-flange.
Holding it in place, re-tighten the two M5 screws to secure it in this unlocked position.
6
Repeat steps
2.4
and
2.5
for the other shipping support.
7
Replace the exterior side covers (Refer to step 1.1).
1
WARNING! CONNECT ALL POWER, COMMUNICATION AND FLUID LINES
PICTURED BELOW BEFORE POWERING ON SYSTEMS AND COMPRESSOR.
If dry
nitrogen is connected, verify that the pressure is set to 15 PSI.
2
WARNING! THE GREEN ROCKER SWITCH ON THE FRONT OF THE COMPRESSOR
SHOULD REMAIN OFF
( I |
O
).
3
Toggle the 20A Breaker switch on the front panel of the compressor ON (
I
| O ).
4
Initialize power to the vacuum control unit by toggling the power switch on the
back of the unit ON (
I
| O ).
5
Initialize power to the system control unit by toggling the power switch on the
back of the unit ON (
I
| O ).
6
Turn on the system control unit by pressing the power button on the front of the
unit (Power button will glow when the unit is on).