W Infusion monitor, Quick guide
1. Starting and using Monidrop device
a) Start the device by pressing the power button.
b) Attach Monidrop to the drip chamber (see instruction in section 2). Adjust the desired infu-
sion rate with the roller clamp.
Shut down the device by pressing the power button
for three (3) seconds. When the de-
vice is switched off, any previously set data and settings will be reset.
Monidrop displays the infusion time, infusion rate (ml/h) set and the volume of fluid (ml) ad-
ministered during therapy.
2. Attaching Monidrop to the drip chamber
a. Open the device retaining clip by
pressing the locking lever and pulling
the clip to release.
b. Position the drip chamber
appropriately. Note the correct
positioning of different types of drip
chambers.
c. Turn the retaining clip to its original
position until it clicks.
3. Adjusting the infusion rate and monitoring the therapy
Adjust the infusion rate by means of the roller clamp on the infusion set to the appropriate
precision. A green LED light and a drop icon on the screen indicate that IV administration is
ongoing.
For additional information on the use and features of the device, the Monidrop
®
W user manual can be downloaded from www.monidor.com
Monidor Oy Elektroniikkatie 3 90590 Oulu Tel. +358 10 3115 860 www.monidor.com
!
Monidrop DOES NOT ADJUST the infusion rate. ALWAYS use a roller clamp to adjust the infusion rate.
Do not fully open the roller clamp.
The Monidrop infusion monitor is intended for measuring IV fluid administration in infusion therapy. The device displays the
infusion rate, total volume administered and treatment duration.
a. Audio notifications can be muted and activated using the no-
tification button.
At the top of the screen, the mute symbol is
displayed when the message is less than 6ml / h. Notifications
with an infusion rate less than 6 ml/h are always muted by default
when the device is turned on.
b. By pressing the audio notification button the audible notifications
are activated if the infusion rate is less than 6 ml/h.
4. Managing audio notifications
5. Screen view
During treatment, the rate and total volume of infusion
can be monitored on the device screen. The functions
are explained below.
Duration of treatment in hours and
minutes (h, min)
Ongoing flow rate (ml / h)
Fluid volume administered (ml)