Page 31 of 36
Electromagnetic compatibility
Electromagnetic emissions
The
Liposat
®
power
is suitable for use in the stated electromagnetic environment.
Customers and/or operators of the
Liposat
®
power
should ensure that they use the
Liposat
®
power
in such an environment.
Measurement of emitted
interference
Level of
conformity
Electromagnetic environment guidelines
High-frequency radiated
interference as per CISPR 11
Group 1
To satisfy its intended function, the
Liposat
®
power
must emit electromagnetic energy. Electronic devices in
the vicinity could be influenced.
High-frequency line-conducted
interference as per CISPR 11
Class B
For areas of application, see chapter „
Suitable
operating environments
”.
Harmonic emissions acc. to
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions acc. to IEC 61000-3-3
Complies
Electromagnetic immunity
Immunity test /
standard
IEC 60601 - testing
level
Compliance level
Electromagnetic
environment guidelines
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
discharge
±15 kV air discharge
±8 kV contact
discharge
±15 kV air discharge
Floors should be made of
wood or concrete or fitted
with ceramic tiles. If the floor
is provided with a synthetic
material, relative humidity
must be at least 30%.
Rapid transient
electrical
disturbances/bursts
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input and
output lines
±2 kV for power supply
lines
±1 kV for input
andoutput lines
The quality of the supply
voltage should be
comparable to that for a
typical shop or hospital
environment.