Appendix
Page 58 of 68
9.3.2
Docon Seal
Immunity test
IEC 60601-
test level
Compliance
level
Electromagnetic environ-
ment - guidelines
Electrostatic dis-
charge (ESD)
IEC61000-4-2
±8 kV con-
tact dis-
charge
±8 kV contact
discharge
Floors should be wood, con-
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humid-
ity should be at least 30%.
±15 kV air
discharge
±15 kV air dis-
charge
Surge voltages
(surges)
IEC 61000-4-5
±1 kV differ-
ential mode
±1 kV differen-
tial mode
The quality of the supply
voltage should be compara-
ble to that for a typical shop
or hospital environment.
Surge voltages
(surges)
IEC 61000-4-5
±1 kV differ-
ential mode
±1 kV differen-
tial mode
The quality of the supply
voltage should be compara-
ble to that for a typical shop
or hospital environment.
Voltage dips,
short interrup-
tions and volt-
age variations
< 5 % U
T
(> 95% dip
in U
T
)
for 0.5 cycle
40 % U
T
< 5 % U
T
(> 95% dip in
U
T
)
for 0.5 cycle
40 % U
T
The quality of the supply
voltage should be compara-
ble to that for a typical shop
or hospital environment. If