Installation and Setup of StakMax® system
34
0112-0133 Rev B
Starting the Software Calibration
To calibrate the software
1.
Double-click the SoftMax Pro icon located on the desktop to open
the SoftMax Pro software.
2.
Click the StakMax icon located on the menu bar. After a few
seconds, the StakMax software interface appears.
3.
Click
Calibrate
.
Note:
Before proceeding with this section, make sure that you have both
pieces of the calibration alignment tool (clear and black), the included Phillips
screwdriver, and the included microplate. You can find these components in
the accessory kit.
Note:
You must have SoftMax Pro software version 5.1 or later to
operate the StakMax instrument.
Note:
Do not use the purple plate adapter when performing the
calibration. Use the purple adapter only when reading a plate in the top
read mode.