Maintenance
SkanWasher 400 User Guide
29
Step 4
Fill a reservoir bottle with distilled water.
Step 5
Connect tubing to instrument.
Connect rinse inlet and pressure tubing to new reservoir bottle.
Verify waste tubing is connected to outlet port.
Step 6
Perform a forward flush to remove the particles.
Press the
[Rinse]
key to enter the Rinse/Prime mode.
Press the
[Start]
key a second time to start the prime program.
(Water will flush out through the holes in the back of the wash
head.)
Repeat if indicated.
Step 7
Reseat inlet channel seal block in the wash head.
Step 8
Run a wash program to verify that particles are removed (see Opera-
tion, Step 5).
Step 9
If particles not removed, and one probe can be identified to be
clogged, replace probe.
Or
If particles not removed, please consult Technical Support to have
inlet channel replaced.
Figure 10: Wash head inlet channel seal block