Troubleshooting
SkanWasher 300 User Guide
43
Discharge test for vacuum pump
Step 1
Turn vacuum pump on with all tubing attached to instrument. Run
pump between 50-150 mbar.
Step 2
Disconnect vacuum tubing from washer.
Step 3
Observe reading on vacuum pump manometer. Reading should drop
to 950-1000 mbar in less than 10 seconds.
Step 4
If test failed, check for obstruction.
Compare pump with internal vacuum sensor
Step 1
Go to Setup Menu/Edit Options/Read Sensor
Step 2
Record the internal vacuum sensor reading
Step 3
Record the pump gauge reading
Step 4
Compare two readings. If the readings are not the same, check for
leaks or blockage in the vacuum system.