Troubleshooting
40
SkanWasher 300 User Guide
Clogged aspiration
probes
z
Use wire probe cleaner.
z
Soak probes overnight in DI water.
z
Consult
Clogged Probe
procedure.
Uneven dispenses
z
Perform
Clogged Probe
procedure.
z
Increase air pressure.
Dripping from dis-
pense probes
z
Check for grime on probes. Perform
Clogged
Probe
procedure.
z
Check wash head. Prime system.
Programs not
retained in instrument
memory
Battery on main board may be low: battery life is
from three to five years. Test the battery strength
to see if it needs to be replaced. Strength should be
3.68 volts ± 5%. Replace battery with Lithium bat-
tery ER6V / 3 6V. Replacement should be per-
formed by a qualified service representative.
Pump continues to
run
Remove tubing from side of washer at pressure fit-
ting. If pump stops, the problem is probably the
tubing caps. If the pump continues running, it is
most likely another internal problem. Call Techni-
cal Support.
Table 2: Troubleshooting Chart
Problem
Possible Remedies
Summary of Contents for SkanWasher 300
Page 1: ...SkanWasher 300 User Guide AN ALL INCLUSIVE AUTOMATIC MICROPLATE WASHER...
Page 10: ......
Page 16: ......
Page 22: ......
Page 30: ......
Page 40: ...Maintenance 30 SkanWasher 300 User Guide Figure 11 Wash Head Inlet Channel Seal Block...
Page 48: ...Maintenance 38 SkanWasher 300 User Guide Figure 16 Row Angle Adjustment Lift Version B...
Page 56: ......
Page 62: ......
Page 64: ......
Page 66: ......
Page 68: ......