Installation
SkanWasher 300 User Guide
6
Connect the tubing from the wash buffer (blue) and rinse reser-
voir (green) to the
Inlet
and
Rinse
ports on the left side of the
instrument.
Figure 2: Wash and rinse liquid reservoirs
Step 6
Insert microplate lift.
Select the appropriate plate type (round or flat-bottom wells) in
use by selecting the correct printed pattern on the side of the
microplate holder (See Figure 3).
Push the plate holder onto the two rods on the microplate lift until
it snaps in place. (see Figure 4).
!! IMPORTANT:
Do not fill the
reservoir bottles completely. Allow
approximately 20% free space for
air pressure.
Waste Reservoir
Vacuum Pump
Washhead
Inlet
Rinse
Hyd. Filter
V
acuum
Pressure
Rinse
Pressure
Inlet