Chapter 8: Selecting Instrument Set Up Options
5032022 C
115
Selecting the Installed Manifold
This option is very important to let the software know the size of the manifold that is
installed on the instrument. A procedure will not run if a different manifold is defined here
than the size defined for the procedure.
There are three manifolds available for the Mu Microplate Washer.
The standard 8-channel manifold works with 96-well plates in portrait orientation.
The optional 12-channel manifold works with 96-well plates in landscape orientation.
The optional 16-channel manifold works with 384-well plates in portrait orientation.
Figure 8-3: The Manifold option
Button
Action
On the left
Press this button to return to
Language
. See
Selecting the Language for the Display
On the right
Press this button to return to
Setup
. See
Selecting Instrument Set Up Options on page
Press this button to scroll forward through the list of available manifolds:
8Needle
,
12Needle
, or
16Needle
Press this button to scroll backward through the list of available manifolds:
16Needle
,
12Needle
, or
8Needle
Press this button to select the manifold and then define the number of wash bottles
connected to the instrument. See
Selecting the Number of Installed Wash Bottles on
.
Table 8-3: Actions of Control Panel Buttons
Summary of Contents for MultiWash+
Page 1: ...5032022 C August 2019 MultiWash Microplate Washer User Guide ...
Page 20: ...MicroWash Microplate Washer User Guide 20 5032022 C ...
Page 50: ...MicroWash Microplate Washer User Guide 50 5032022 C ...
Page 64: ...MicroWash Microplate Washer User Guide 64 5032022 C ...
Page 144: ...MicroWash Microplate Washer User Guide 144 5032022 C ...