Environment Control Option Operations
5014166 D
15
To set the temperature
1.
On the front of the Environmental Control Option Controller or
the Systems Power & Options Controller, view the current
sample plate temperature.
Figure 1-7:
Sample Plate Temperature control interface
2.
To view the temperature set point, press
.
3.
To increase the temperature set point, press
and
.
4.
To decrease the temperature set point, press
and
.
After you release
, the current temperature in the chamber is
displayed.
Setting Up the Water Reservoir
Air supplied to the environmental enclosure flows through the water
reservoir to make sure the air has a high degree of humidity.
Tools required for setup:
Syringe
Deionized Water (preferably sterilized)
Note:
For the ImageXpress Micro Confocal System, you can set the
temperature with the Instrument power switch in the On position. To
regulate the temperature in the chamber, the Sample Heater power
switch must be also in the On position.