FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H
165
Choosing Analysis Options
In the
Analysis
process, click on the
Grouping
,
Correction
or
Export
buttons to define the analysis parameters to be applied to the
experimental data. For more information about these options see
When Finished Creating the Protocol
When you have finished defining the experiment parameters of the
processes in the Experiment window, click
Save As
and save the file as
an *
.fmp
file.
Optimizing the Optics and Fluid Dispensing
Optimizing the Hardware Settings
Prior to screening, the following hardware components should be
optimized for signal output and detection:
•
Excitation intensity
•
Exposure time
•
Number of reads
•
Camera gain for EMCCD camera
•
Gate Open for ICCD camera
•
Pipettor height
•
Dispense speed
•
Dispense volume
•
Cell parameters for suspension assays.
Optimizing Optics Hardware
Optimization is available to help amplify weak signals or reduce
saturation. Modifying read settings may not alter the quality of your
data because it only regulates the amount of light emitted or detected
by the system. For instance, doubling the excitation intensity will
double the RLUs detected by the system, but your signal-to-noise ratio
and Z'-factor may remain the same.
To optimize the fluorescence signal:
1.
Place the cell plates on the read positions.
2.
Start with the excitation intensity, camera gain and exposure
time settings shown below. These settings are the most
frequently used to measure basal fluorescence signal.
Summary of Contents for FLIPR Tetra
Page 1: ...FLIPR Tetra High Throughput Cellular Screening System User Guide 0112 0109 H December 2011...
Page 12: ...Contents 12 0112 0109 H...
Page 16: ...System Overview 16 0112 0109 H...
Page 40: ...System Hardware Features 40 0112 0109 H...
Page 148: ...Exchanging Hardware 148 0112 0109 H...
Page 156: ...Calibration and Signal Test 156 0112 0109 H...
Page 196: ...Running an Experiment 196 0112 0109 H...
Page 232: ...Robotic Integration 232 0112 0109 H The following drawings illustrate these requirements...
Page 282: ...Data Processing Algorithms 282 0112 0109 H...
Page 294: ...Consumables and Accessories 294 0112 0109 H...
Page 298: ...Using AquaMax Sterilant 298 0112 0109 H...
Page 302: ...Electromagnetic Compatibility EMC 302 0112 0109 H...