Exchanging Hardware
142
0112-0109 H
Uninstalling a Filter
To uninstall a filter:
1.
Open the lower front door of the FLIPR
®
Tetra System.
2.
Remove the
Filter Cover
, located below the left LED module.
3.
Using the filter tab that protrudes slightly from the filter tray,
push the filter up out of its position and remove it from the
instrument. Make sure only to grasp the filter by the tab. DO
NOT touch the filter’s surface.
4.
Repeat Step 3 for the remaining emission filters you want to
remove.
Note:
Once the lower front door is open, system power is
disengaged.