3 Operating Procedures
3.5.6 Assay Plate Type
You can select the type of assay plate you want to use from the following menu:
Assay Plate Types
Be sure to select a plate type that matches the type and well configuration of the actual
microplate you will be using. Plates do vary, particularly with regard to well bottom
height.
Note:
If you change the Assay Plate Type setting, within a Plate section, the well
assignments stored in the previous template will be discarded. The previously created
groups remain, however, so you can select new wells and apply existing groups to them.
Note:
The Assay Plate setting takes precedence over all other fluidics module settings
and will affect the correctness of other settings.
Important!
If you go back and change this setting after you have selected settings that
follow, for example, Pipette Tips Layout, Compound Source, or Compound and Tip
Columns, the earlier settings may be automatically reconfigured to reflect the new Assay
Plate Type and well layout. Be sure to check any earlier assignments to ensure they
remain correct.