Troubleshooting
0112-0151 K
99
Dispense Issues
Table 6-5:
Dispense Issues
Problem
Possible Causes & Remedies
No liquid in the plate after
dispense.
Verify that the dispense parameters (inlet,
volume) are set correctly.
Verify that all appropriate bottles and tubing are
connected to the instrument, and that the bottles
are not empty.
Verify that the system has been primed properly.
Dispense precision is out of
spec (high CV, left-right
bias, empty wells, or
outliers).
Air bubbles in the wash head or system fluidics:
•
Run a full Prime, such as the One-touch Prime
program.
One or more dispense probes might be clogged:
•
Follow the procedures in
to unclog the probes.
Dispense accuracy is out of
spec (high/low dispense
volume).
Dispense adjustment values might be incorrect:
•
Adjust the dispense accuracy, see
dispense accuracy for a program on page 68
.
Dripping from the dispense
probes.
Air bubbles in the wash head or system fluidics:
•
Run a thorough prime program (such as the
one-touch Prime program).
One or more dispense probes might be clogged:
•
Follow the procedures in
to unclog the probes.
Dispense overfills the
plate.
Dispense volume might be set too high:
•
Verify that the total dispense volume in the
program does not exceed the maximum well
volume for the plate in use.
Dispense adjustment values might be incorrect:
•
Adjust the dispense accuracy, see
dispense accuracy for a program on page 68
.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...