System Fluidics Issues
Table 6-2:
System Fluidics Issues
Problem
Possible Causes & Remedies
Air in the bottle lines.
Prime lines using the one-touch Prime
program.
Leaks from the liquid inlet
quick connects on the left side
of the washer.
When the bottle tubing is disconnected from
the quick-connect, it is normal for a small
amount of liquid to leak from the inlets.
The prime trough does not
drain and is filled with liquid.
Manually purge the trough by touching Purge
in the Options tab. This will also empty the
wash head.
Liquid under instrument.
If the overflow drain (see
wet, and liquid is overflowing from the
prime/soak trough:
•
Verify that the total dispense volume in
the program does not exceed the
maximum well volume for the plate in
use.
•
Adjust the dispense accuracy, see
adjust dispense accuracy for a program on
page 68
•
Remove any blockage in the prime/soak
trough.
If the overflow drain is completely dry:
•
Check around the instrument for other
sources of leaks, such as the tubing
connectors from the bottles.
•
If the leakage is coming from within the
instrument, contact Technical Support.
Note:
A small amount of liquid will not
cause damage to the instrument. If the
trough is full, manually purge the trough
by touching Purge in the Options tab.
Carefully wipe up any spills under the
instrument.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...