AquaMax
®
2000/4000 Microplate Washer User Guide
90
0112-0151 K
To recover the microplate and wash head manually
The Reset button on the bottom right corner of the Options tab performs
a software reset without power cycling the instrument. If the washer fails
during an operation, perform a reset to manually retrieve any microplate
still stuck in the system.
1.
To return the wash head and microplate to their unloaded
positions, touch
Reset
.
2.
Retrieve the microplate.
If the plate carriage cannot be unloaded by a software Reset,
continue with the following steps.
3.
Lift the wash head gently by the finger guard until it clears the
microplate.
4.
Disengage the lever mechanism and remove the wash head.
5.
Carefully slide the plate carriage out.
6.
If the plate carriage is stuck, touch
Load/Eject
in the
Options
tab.
7.
Remove the plate from the plate carriage.
CAUTION!
Do not touch or put pressure on the dispense or
aspiration probes, as they can be damaged easily.