Getting Started
0112-0151 K
35
To run a one-touch maintenance program
1.
Make sure that all appropriate reagent bottles are connected to
Liquid Input and Pressure fittings on the left side of the
instrument.
2.
Make sure that the waste tubing is connected to the Waste
Output fitting and the Waste bottle.
3.
On the
Main
tab, touch the appropriate one-touch program
button.
When a program is running, the touch screen displays a status
dialog. (See
Stopping Programs
To interrupt a program in non-emergency conditions
When you run a program, the touch screen displays a status dialog. (See
•
For non-emergency stops touch
Cancel
.
Cancel stops the program when appropriate. It turns off the
pumps, resets the wash head, moves the plate carriage to the
plate load position, and performs program clean up before
returning control to the touch screen.
To interrupt a program in emergency conditions
When you run a program, the touch screen displays a status dialog. (See
1.
For emergency stops touch
Stop
.
Stop halts the program immediately. It turns off all pumps and
motors, closes valves, and releases system pressure.
2.
Touch
OK
to recover and remove the plate.
A software reset is performed before returning control to the
touch screen.
Note:
For information about one-touch program defaults, see
. One-touch programs can be copied and
edited.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...