AquaMax
®
2000/4000 Microplate Washer User Guide
154
0112-0151 K
Error Recovery
Errors ³ 600 are caused by improper input. No recovery is required. The
error prevented instrument operation. Instrument operation continues
with the next valid command.
Errors between 100 and 199 permit continued operations. Continued
operation is possible following acknowledgement and/or correction of
the error. For example: empty waste container to correct waste full error.
Errors between 200 and 299 or less than 100 are Fatal Errors: only status
commands are available. A reset is required for continued operation.
Instrument Response
The host computer should receive a response from a directly connected
instrument within 1 second. If no response is received within the allowed
time, the host computer should generate a timing error and can assume
that the microplate washer is not connected or not functioning properly.