Wash Issues
Table 6-7:
Wash Issues
Problem
Possible Causes & Remedies
Some wells overflow during
wash.
Aspiration probe tips might be touching the
well bottoms during bottom wash:
•
Verify that all plate parameters in the
plate library match the plate in use. If
necessary, re-calibrate the well depth
using the Auto feature according to
Wash solution surface tension causing uneven
aspiration:
•
Add surfactant (such as Tween) to the
wash solution.
Wash volume is too high:
•
Decrease the wash volume. The wash
volume should not exceed 10x well
volume.
One or more aspiration/dispense probes might
be clogged:
•
Follow the procedures in
to unclog the probes.
Liquid left in wells after Wash
program step.
By design, the Wash step leaves some residual
liquids in the wells. Follow the Wash step with
an Aspirate step to remove the liquids if
desired.
Summary of Contents for AquaMax 2000
Page 1: ...www moleculardevices com AquaMax 2000 4000 Microplate Washer User Guide 0112 0151 K June 2014...
Page 8: ...AquaMax 2000 4000 Microplate Washer User Guide 8 0112 0151K...
Page 30: ...AquaMax 2000 4000 Microplate Washer User Guide 30 0112 0151K...
Page 82: ...AquaMax 2000 4000 Microplate Washer User Guide 82 0112 0151K...
Page 110: ...AquaMax 2000 4000 Microplate Washer User Guide 110 0112 0151K...
Page 112: ...AquaMax 2000 4000 Microplate Washer User Guide 112 0112 0151K...
Page 114: ...AquaMax 2000 4000 Microplate Washer User Guide 114 0112 0151K...