3. Operating Instructions
3.11.26. Remote Control
Using the accompanying CT Viewer™ 2 software, the CT100B can be operated by remote
control. The remote control system allows the operator to manipulate traces and store movies of
traces over time. Remote control only functions over Ethernet connections. It does not work
when the CT100B is connected by USB only.
Refer to Section 4.1.3 Send Saved Traces Using Ethernet and Section 4.1.4 Using Remote
Control for instructions on connecting CT Viewer 2 to the CT100B and enabling remote control.
Please consult the
CT Viewer 2 Quick User Guide
for more information about using remote
control and manipulating a remote TDR trace.
3.11.27. Cable Type Library
The CT100B has a library of common cable types and associated velocities of propagation. Vp
for these cable types can be loaded from the library. The user can add custom cable types to the
CT100B as well. The user-added cable types are stored and accessed separately from the
standard cable types.
3.11.27.1. Load a Cable Type’s Vp
To load a cable type, follow this procedure:
1. Press the
FILE
button. The
File
menu appears.
2. Select
Reference Cable Types
or
Custom Cable Types
from the menu. An interface appears
that allows you to scroll through the cable types with the
M-FUNCTION
knob.
3. Highlight the desired cable type and press
Select
. Vp is now set to the value for that cable
type.
3.11.27.2. Custom Cable Types
The CT100B is able to save custom cable types. Use the following procedure to save a custom
cable type:
1. Press the
FILE
button. The
File
menu appears.
2. Select the
Custom Cable Types
menu option to access user-editable cable types with menu
options
Add New
and
Select
.
3. When
Add New
is chosen, a window appears for entering a cable type name, a Vp for
velocity of propagation, and an expected characteristic impedance for the cable. When the
window first appears, Vp will be set to the current system Vp, while impedance defaults to
50 ohms.
3.11.28. Other Measurement Settings
3.11.28.1. Change Horizontal Units
Horizontal distance measurements can be displayed in meters, inches, feet, yards, or
centimeters.
1. Select the
MENU
→
Settings
→
Meas. Settings
menu.
60
CT100B TDR Cable Analyzers Operator’s Manual