EN
OPERATOR'S MANUAL
65
APPENDIX -
MAINTENANCE
In addition to correct use, the user needs to perform ordinary maintenance in order to
guarantee safe, efficient operation over the device’s entire life.
FOREWORD
For better quality of maintenance, supplement routine checks with regular periodic check-ups
that can be performed by Technical Service Department (
see Appendix
).
It is also fundamental to perform a periodic thermo-disinfector validation, i.e. a check of
process parameters and their comparison with the reference values.
Refer to ‘Thermo-disinfector periodic validation' in the next part of
this Appendix.
The ordinary maintenance described below consists in easy manual operations and
preventive interventions involving simple instruments.
WARNING
IN THE EVENT OF REPLACEMENT OF COMPONENTS OR PARTS OF
THE DEVICE REQUIRE AND/OR USE ORIGINAL SPARE PARTS ONLY.
ORDINARY
MAINTENANCE
PROGRAMME
The table summarizes the maintenance interventions required to maintain the thermo-
disinfector in good working order.
In case of heavy use, we recommend to shorten maintenance intervals:
DAILY
Clean the filter at the bottom of the chamber
Clean external surfaces
Clean chamber surface from possible residues of detergent
Clean residues of water or salt from the surface when filling with salt
WEEKLY
Clean the washing chamber and its accessories
Disinfect external surfaces
PERIODIC
See Scheduled Maintenance messages
ANNUALLY
Thermo-disinfector validation (
see dedicated paragraph
)