System Two LED Installation and Service Manual (1003385 Rev. C)
Page 61
Section 6
Mounting the securing segments
Figure 38: Mounting the securing segment
Figure 39: Mounting the socket covers and decorative
caps to the Extension Arm
Figure shows a simplified illustration of the
Extension Arm [4] without cables, Spring Arm and
end device.
1. Lubricate the securing segment
②
with white grease.
2. Lubricate the Spring Arm pivot (not illustrated in the
Figure) with white grease through the fitting aperture
of the securing segment
②
.
3. Route the securing segment
②
into the Extension
Arm until it snaps into place.
The spring-actuated latch pin
①
audibly clicks into
place.
4. Check securing segment
②
is securely in place;
The latch pin
①
must audibly click into place.
The latch pin
①
is at the same height as its opposite
latch pin.
5. Dismantle and check the second securing segment.
The second securing segment is on the opposite
side.
6. Re-mount the second securing segment
Mounting the socket covers and decorative caps to
the Extension Arm
1. Screw the left-hand socket cover with optional cover
plate
②
onto the Extension Arm [4] using 2 cross
recessed head screws
①
.
2. Place the right-hand cover panel with optional cover
plate onto the left-hand socket cover
②
such that all
the locking pins
③
engage with each other and
screw it onto the Extension Arm [4] using 2 cross
recessed head screws
①
.
3. Check that the socket cover with optional cover
plates
②
is securely in place;
The socket covers with optional cover plates
②
must
tightly engage with each other without gaps and be
tightened with 2 cross recessed head screws
①
.
Mounting the decorative caps
*** Notice ***
Damage to the decorative caps
To prevent damage to the decorative caps;
The mounting steps described below must be
executed in the correct order:
1. Attach the locking mechanisms
⑥
of the decorative
caps
④
into the 2 fittings
⑤
on the covers
②
.
2. Fold the decorative caps
④
towards the Extension
Arm 4 and insert the straps
⑧
into the cutout
⑦
on
the socket covers
②
.
3. Check that the decorative caps
④
are securely in
place;
The decorative caps
④
must securely engage in the
socket covers
②
without gaps.