Section 6
Handle Sterilization and Cleaning
Figure 35: Handle Sterilization
Cleaning Agents
The System Two LED features a removable
sterilizable Handle.
Handle Sterilization
Remove the sterilizable handle by pressing the
Handle Button near the base of the handle towards
the center and pull the handle off the handle post.
Sterilize the handle utilizing steam sterilization of
minimum 250
Fahrenheit for a minimum of 30
minutes in compliance with AAMI-SSSa-1988 Good
Hospital Practices, Steam Sterilization and Sterility
Assurance, or equivalent method.
Cleaning
Follow the safety instructions on Page 56.
*** Warning ***
Risk of contamination and infection of the patient
Parts of the pendant system and the adaptations
are made of plastic. Solvents can dissolve plastic
materials. Strong acids, bases and agents with an
alcoholic strength of more than 60% can lead to the
plastic materials becoming brittle. Detached
particles can fall into open wounds. If liquid
cleaning agents are allowed to penetrate the
pendant system and the adaptations, excess
cleaning liquid may drip into open wounds.
Recommended cleaning agents Use a mild soap
solution or a regular dishwashing product.
1. Wipe the surfaces of the appliances with a
moderately moist cloth; add a mild soap solution
(dishwashing product) if required.
2. Afterwards, carefully wipe the surfaces dry with a
clean cloth.