System Two LED Installation and Service Manual (1003385 Rev. C)
Page 57
Section 6
Disinfection
Disinfection method
Follow general safety precautions
*** Warning ***
Other than the Sterilizable Handle on the Light Head, the
Arm System is not suitable for sterilization
Avoid damage
Make sure that no liquid penetrates the appliance whilst
cleaning it
To prevent damage to plastic parts, refrain from using
abrasives or alkaline, acidic or corrosive cleaning agents.
Do not use bleaching agents on stainless steel parts.
Deploy trained technical specialists only and abide by
national regulations.
Cleaning/disinfection must be carried out by trained
technical specialists only.
*** Caution ***
Health hazard
Disinfectants can contain substances hazardous to health
which, when in contact with the skin and eyes, can cause
injuries or affect the respiratory organs when inhaled. Observe
the protective measures;
Observe the hygiene regulations.
Adhere to the disinfectant manufacturer's instructions.
Perform surface disinfection every working day and in case
of contamination.
Wiping disinfection is the standardized disinfection method
prescribed for the Light System. Hygiene regulations and
related safety instructions for the disinfection methods to be
applied must be defined by the operator.
In case of contamination with potentially infectious material
(e.g. blood, body secretion or excrement) the surfaces
must be immediately and specifically disinfected.
Make sure you apply the disinfectant in the correct
concentration.
For surface disinfection do not spray, but wipe, the
surfaces.
Wiped surfaces may only be used after the disinfectant has
dried.