971B UniMag™
7
Operation manual
971B Functions
User Switch
I/O Connector
Red or Green LED Status Indicator
User Switch
The user switch turns ON the cold cathode sensor.
Refer to page 18 for operation procedure.
The cold cathode sensor should only be turned on when vacuum pressure is below 5×10
-3
Torr
If the transducer is delivered with customer specified parameters, the User Switch is disabled.
For enabling the switch see page 22.
LED Status Indicator
The red/green LED status indicator has the following stages:
1. GREEN
Normal operation
2. 2 sec. RED
Power on sequence
3. GREEN 1 sec. flash cycle
Test mode TST!ON (see page 23)
4. GREEN 0.5 sec. flash cycle
Cold Cathode high voltage turned on
5. Red 0.5 sec. flash cycle
Cold Cathode high voltage turned on, but ionization has not
started.
6. RED
Sensor failure (see page 34)
7. OFF
Power off