2
Initial Setup Procedures
2-10
Series 370 Stabil-Ion Instruction Manual - 370119
2.1.4.2 Handshaking
Table 2-8
Inputs to 370 Controller
Line
Pin
Description
Internal Switch
Function
Factory
Setting
CTS
&
DSR
5
6
CLEAR to SEND
and DATA SET
READY: When
used, both must be
TRUE in order for
370 to send the
next byte in its
message or data.
CTS=1 and DSR=1:
When ON, forces the
functions TRUE and
thus assumes host is
always ready to
receive.
Both ON
DCD
8
DATA CARRIER
DETECT: Must be
TRUE at the time
each character is
received or that
character will be
ignored by 370.
DCD=1: When ON,
forces DCD function
TRUE so 370 will
receive all characters
sent to it (as long as
RTS is in
de-asserted state).
ON
Figure 2-13
Top View of RS-232 Module
X
OFF
X
ON