background image

4.2. Changing Chosen Wavelengths (metallic type only)

1.  Select “Laser” and enter.

2.  Select the wavelength you wish to use and enter. If you wish to save  

 

this wavelength as the startup default, press save before exiting.

3.  If you wish to modify one of the wavelengths, select the closest  

 

wavelength, press the right navigation key.

4.  Now, using the up/down keys to change the numbers and the right/left  

 

keys to move to the next number, modify to the wavelength you want.

5.  When finished, press “save” and exit. If you only want a temporary  

 

change, exit without “save”.

4.3. To Set Type of Laser Being Used

From main measurement screen, select “Laser” and enter. Choose the 

desired laser type or wavelength. If you want this one to be the startup 

wavelength, press save before exiting. 

4.4. Setting Startup Configuration

1.  Turn on the Nova II with the sensor connected, select “Menu” and  

 

enter. Select “Configure”. 

2.  Now, with the navigation keys select “Startup Mode”, select “power”  

 

or “energy” as the startup mode. Press enter to exit. 

3.  Now, with the navigation keys select  “Display Mode”. Press the enter  

 

key and select the type of screen you wish to start up with. Press  

 

save and exit. If you do not want to save this as the startup mode but  

 

only change for now, exit without save. (This choice is saved in the  

 

instrument and not the sensor so the Nova II will start up from the  

 

screen selected for all sensors).

4.  Now select “Energy Range”, press enter and choose the manual energy  

 

range you wish to be the default.

5.  Now select “Average” to select what period you wish to average power  

 

over or select the default, “none”. 

6.  If the sensor is a diffuser sensor, now select diffuser in or out.

7.  Now select “Max Pulse Len” and choose the shortest pulse length  

 

setting longer than your laser’s pulse length. 

Warning:

  Incorrect readings will result if pulse length is not set up correctly. 

8.  Select “Laser” and select the laser type you wish to be the default.

9.  Press “save” to save all present settings then “exit” to get back to the  

 

main measurement screen.

Note: 

Some sensors have only the long pulse setting for all pulse lengths. In 

that case N/A appears. For further details on configuration see Sections 6.2 

and 6.3 of the manual.

4.5. Energy or Average Power Measurement

Warning:

Do not exceed maximum sensor limits for power, energy, power density and 

energy density as listed in tables 5 and 6 section 10.2 of the manual, Sensor 

Specifications.  Otherwise, there is a risk of damaging the absorber.

With the pyroelectric sensor, you have been supplied a test slide with the 

same coating as on your pyroelectric detector. You can also obtain this slide 

from your dealer.  You should use this slide to test the damage threshold 

with your laser pulses.  If the slide is damaged, then either enlarge your 

beam or lower the laser energy until damage is no longer seen.

4.6. To choose Energy or Average Power Measurement

To switch from the main energy measurement screen to power 

measurement, press “power”. To switch from the main power measurement 

screen to energy measurement, press “energy”.  

Nova II Quick Refer

ence

NOVA II 

Quick Reference

Ophir Optronics Solutions LTD.

Science Based Industry Park P.O.B. 45021, Jerusalem  9145001, Israel

Phone +972-2-5484444

https://www.ophiropt.com/photonics

rev 1.45-5/17.Feb.19/be  Ophir P/N 1J06036

1

 Getting Started

The Nova II is equipped with “soft keys”.  That is, the functions 

of the keys change as indicated by the legend above each key. 

It is also equipped with 4 navigation keys and a round enter key 

in the center.

To connect sensor to the Nova II meter:

Insert the 15 pin D type connector of the measuring sensor 

cable into the socket marked “Head Input“ on the rear panel of 

the Nova II meter.

To switch the Nova II on:

To switch on, briefly press the on/off/backlight switch (leftmost 

key). The unit will switch on, and the display will appear.  Note 

that the sensor must be plugged in before the unit is switched 

on. To switch the backlight on or off, press the on/off/backlight 

switch briefly again. To switch the Nova II off, press the on/off/

backlight switch and hold it for ~ 3 seconds until the display 

blanks. 

If you wish to save the current Nova II configuration, use the 

“configure” function. (See section 2.3.2.).

To set the display modes:

1.   Disconnect the sensor and switch off then on again.  “Head  

 

Disconnected” will appear.

2.   Select “Display Mode” and enter. Choose the desired mode,  

 

press save. The Nova II will now be in this display mode for  

 

all sensors.

3.   Select “Date” and press the enter button. Using the  

 

navigation keys set to the correct date. Press save.

4.  Select “Time” and enter. Using the navigation keys, set to  

 

the correct time. Press save.

5.  For adjusting RS232 baud rate and An. Output voltage see  

 

the main manual.

Reviews: