Preparing for Operation
Series 358 Micro-Ion Controller Instruction Manual - 358013 - Rev. B
91
Op
er
ation
C
a
p
a
cita
nce Ma
nom
eter
Servic
e
Pre
p
a
ring
fo
r
Op
er
atio
n
Command Syntax for
RS-485 Computer
Interface
DG
Definition:
Turn degas ON or OFF.
Modifiers:
ON or OFF.
Response:
OK if command accepted, or INVALID if
rejected.
Example:
From computer:
#AADG ON CR
From controller:
OKCR
• Command is INVALID if the Ion gauge is OFF.
• A response to the DG ON command of OK indicates only that a signal
requesting degas has been sent to the electrometer. Degas may fail to
activate, e.g., if the pressure is above 5 x 10
–5
Torr. Use the DGS
command (see below) to verify that degas has been successfully initiated.
DGS
Definition:
Display degas status.
Modifiers:
None.
Response:
ASCII 1 if degas is ON, 0 if degas is OFF.
Example:
From computer:
#AADGSCR (Spaces may be omitted.)
From controller:
1CR (Indicating degas is ON)
DS
Definition:
Display pressure reading.
Modifiers:
IG or CG1 or CG2
Response:
ASCII string representing the pressure for the
selected gauge.
Example:
From computer:
#AADS CG1 CR
From controller:
1.20E–03CR
• The DS CG1 and DS CG2 commands are used to display the pressures
from the second and third display lines.
• If the ion gauge is turned OFF, or is in its first few seconds of operation,
the controller will return 9.90E+09.
• The DS IG command will return pressure if the gauge is ON, and
9.90E+09 if it is OFF.
Summary of Contents for Granville-Phillips 358 series
Page 2: ......
Page 8: ...Table of Contents 8 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 26: ...Chapter 2 26 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 48: ...Chapter 3 48 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 94: ...Chapter 5 94 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 104: ...104 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 116: ...Chapter 8 116 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 120: ...120 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 121: ......